•Best value professional’s
autoranger
•9 functions/36 ranges
•Oversized, 4000-count
display with bargraph
•True-rms measuring
•Min/Max, Data Hold,
Relative (∆) modes
•Holster included
•CAT II 1000 V, CAT III 600 V
rated
Model 2015
True-rms Digital Multimeter
2015 Specifications
(at 23 °C ± 5 °C, < 75 % R.H. non-condensing)
DC Voltage
Ranges 400 mV, 4 V, 40 V, 400 V, 1000 V DC
Accuracy ±(0.25 % rdg + 2 dgts)
Resolution 100 µV in 400 mV range
Input Impedance 400 mV: >1000 MΩ
4 V: 11 MΩ
40 V to 1000 V: 10 MΩ
AC Voltage
Ranges 400 mV, 4 V, 40 V, 400 V, 750 V AC
Accuracy 400 mV (50-500 Hz): ± (1 % rdg + 5 dgts)
4 V to 750 V (50-500 Hz): ± (0.8 % rdg + 3 dgts)
4 V to 400 V (500-1 kHz): ± (1.0 % rdg + 5 dgts)
750 V (500-1000 Hz): ± (1.2 % rdg + 5 dgts)
Resolution 100 µV in 400 mV range
Conversion True RMS
Crest Factor 3:1
Input Impedance 400 mV: > 1000 MΩ
4 V: 11 MΩ
40 V to 750 V: 10 MΩ
DC Current
Ranges 4 mA, 40 mA, 400 mA, 20 A
Accuracy 4 mA to 400 mA: ± (0.8 % rdg + 2 dgts)
20 A: ± (1.2 % rdg + 4 dgts)
Resolution 1µA in 4mA range
Burden Voltage 4 mA: 100 mV/mA
40 mA: 10 mV/mA
400 mA: 1.7 mV/mA
20 A: 30 mV/A
AC Current (50 Hz-1000 Hz)
Ranges 4 mA, 40 mA, 400 mA, 20 A
Accuracy 4 mA to 400 mA: ± (1.2 % rdg + 4 dgts)
20 A: ± (1.5 % rdg + 5 dgts)
Resolution 1 µA in 4 mA range
Conversion Type True-rms
Crest Factor 3:1
Burden Voltage see DC Current
Resistance
Ranges 400 Ω, 4 kΩ, 40 kΩ, 400 kΩ, 4 MΩ, 40 MΩ
Accuracy 400 Ω: ± (0.8 % rdg + 4 dgts)
4 k-400 kΩ: ± (0.7 % rdg + 2 dgts)
4 MΩ: ± (1.0 % rdg + 4 dgts)
40 MΩ: ± (1.5 % rdg + 4 dgts)
Resolution 100 mΩin 400 Ωrange
Test Voltage All Ranges 0.4 V DC
Diode Test
Range 0 to 4 V
Accuracy ±(1.0 % rdg + 1 dgt)
Test Current 0.6 mA
Resolution 1 mV
Display Forward junction voltage
Open Circuit Voltage 3.2 V DC
Continuity Test
Range 400 Ω
Audible Threshold < 40 Ω
Response Time Approx. 100 msec
Open Circuit Voltage 400 mV dc
Measuring Digital Multimeter